onductivity cell problems and solutions

Publisher:admin    Release time:2017-03-03 11:31

Phenomenon: Baseline is unstable and noise is large
Possible reasons: (1) the inside of the conductivity cell is not smooth and burr.
(2) after the conductivity of the metal part of the metal and the host circuit poor contact.
Solution: (1) to remove the burr needle.
(2) with solder wire click on the metal part of the conductive plate.